Compur Ex plus Gas Detector User Manual Page 26

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 25
22
Maintenance and Calibration Triple Plus+ and Triple Plus+ IR
VII. Maintenance and Calibration
Crowcon recommends monthly gas response checks and a recalibration interval
of 6 months. Flammable sensors will respond to hydrogen and most hydrocar-
bons but will only be calibrated for a specific gas, typically methane.
Note:
when the instrument is placed into a charger or charger/interface which is fit-
ted with the new LIBRA lithium-ion battery pack the unit will turn off and start charg-
ing. If performing calibration or maintenance work, turn the instrument on when
fitted into the charger or back on again if switched off when placed in the charger.
7.1 Recalibrating
Read about the ‘CALIBRATE?’ option in SETTING UP (CALIB) and the CALIBRATE
functions in Portables PC software as either method may be used to adjust gain
values, but a knowledge of both is helpful.
Zero the unit in clean air. Locate the aspirator flowhead under the ledge above
the sensor panel and secure with the half-turn fasteners.
Starting with flammable gases and moving on to toxic, connect the appropriate
cylinder of test gas to the inlet of the flow head via a flow meter. Direction of
flow is marked on the flowhead. Adjust the control valve on the flow meter to
achieve a flowrate of 0.5 litre/min.
Wait until the reading has stabilised on the Triple Plus+/Plus+ IR display before
programming in the new level. Now shut off the valve and disconnect the
supply of test gas. Aspirate fresh air through the flowhead using the rubber
bulb until the reading has returned to zero. If there are signs of cross sensitivity
between two or more sensors, wait for all readings to settle before proceeding
with calibration of other channels.
For units with both %LEL (pellistor) and %Volume (flammable sensors) see note
at end of ‘CALIB’ Section 3.2 (p13).
7.2 Replacing the Lead-Acid Battery Pack
To change the battery pack, ensure you are in a safe, non-hazardous area.
Unscrew the tamperproof screws which retain the battery hatch using a 2 mm
Hex. key. Remove the hatch and tip the battery pack out. Disconnect the spade
terminals. When refitting, be sure to observe the correct polarity; red wire to
positive, black to negative. Replace the battery pack and hatch and secure with
the tamperproof screw. Crowcon recommends fitting the new Li-ion battery
pack when replacing a lead-acid battery that has reached the end of its useful
life. Alternative batteries, other than the Crowcon Li-ion battery pack, should not
be used because they will invalidate approvals and may impair performance.
Page view 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 39 40

Comments to this Manuals

No comments