Compur Statox 501 Control Module
One for all
Statox 501 HRC and ARE
Two Sensor Heads covering all explosive Hazards
Statox 501 sensor heads HRC and ARE
detect combustible gases in the range
0 100 % LEL. Both operate with catalytic
sensors. This sensor type burns the gas
to be detected on the catalytic surface
of the sensing element and thus increases
its internal temperature and electrical
resistance. This parameter is easy to
measure. Therefore these sensor heads
are robust and affordable.
Catalytic sensors respond to any kind of
combustibles although not with the same
sensitivity for different substances. Statox
501 HRC is the better choice for volatile
hydro carbons, while the ARE type is very
good for solvents and fuels.
Both sensor heads operate in conjunction
with the Statox 501 control module which
provides the power supply and processes
the signal. It features three powerful relays
and a 4 20 mA analog output.
The combination sensor head / control
module constitutes a complete gas detection
system for combustible gases. There is one
potential disadvantage of catalytic sensors:
Their reliability may be compromised by
catalyst poison or oxygen defi ciency. In this
case one of Compur Monitors infrared gas
detectors will be the better alternative.