Test Gas Generator for easy Bump Test
of Gas Detectors
Safety in ten Seconds
Personnel working in chemical plants, uses gas
detectors to protect themselves from exposure to toxic
gases. Gas leaks fortunately happen very rarely, so the
gas detector will normally read “zero” all the time.
The “zero” reading may have two reasons:
a) The gas concentration is zero.
b) The gas detector is defective.
For this good reason a daily performance test of the
gas detector is mandatory in many plants. A complete
calibration would be too time-consuming, but a bump
test will prove that the detector is working. It will reliably
indicate any damage by wear and tear or abuse.
The gas generator can be operated by everyone without
extra training. The gas generation starts, as soon as a
gas detector is attached to it. It stops gas generation
automatically after ten seconds, or if the gas detector
is removed. If the detector fails to alarm within this time
interval, it needs to be serviced.