5.1 Using the diagnostic box
Switch the diagnostic box on. Then connect it to the opto- coupler receptacle on sensor head bottom.
Keep the LED`s and the phototransistors of the opto- coupler clean for proper data transmission.
5.1.1 Operation and error codes of the diagnostic box
Note: If a combination of push button operations is requested ( f. i. ST + ENT), the buttons must be
operated in a short time distance.
Example: ST + ENT initiates a self test. The display says „WAIT-TEST RUN“ until the self test is
finished.
Diagnostic box queries:
WAIT; after a while reading of actual measured value in ppm
Sensor head measuring range
Status passed (System OK) or error message (see 5.1.2)
(initiates self test), TEST Run
ZERO RUN (start zero adjustment)
* only available with systems with zero adjustment
Diagnostic box error messages:
Faulty or too slow input. Repeat.
Calibration gas concentration out of range.
Calibration factor is out of range.
Measuring range or gas not correctly programmed.
Time error during data transmission.
Faulty data transmission.
No self test due to too low temperature.
Temperature out of the admissible range.
Temperature is back in the admissible range.
Charge calibration box battery.
Zero adjustment not possible. Wrong EPROM in sensor
head and control module.
Sensor is flushed with air because the gas concentration is
too high.
Calibration routine has been cancelled with CE button.
Calibration not possible within the time limit.